Experimental browser for the Atmosphere
Probe arcing ( u-of-o-nmr-facility.blogspot.com/2009/11/prob... ) may damage the electronics in an NMR probe (or RF amplifiers) if allowed to happen over extended periods of time. It would not cause a magnet quench.
Apr 23, 2025, 12:35 AM
{ "uri": "at://did:plc:stc2xba5cokpts4oho7adbef/app.bsky.feed.post/3lngx5dntjs2e", "cid": "bafyreihj6kdaczpg2wqvktezdxkimvjj2p7ks3kd3vlk46qdur4kzngtyy", "value": { "text": "Probe arcing ( u-of-o-nmr-facility.blogspot.com/2009/11/prob... ) may damage the electronics in an NMR probe (or RF amplifiers) if allowed to happen over extended periods of time. It would not cause a magnet quench.", "$type": "app.bsky.feed.post", "langs": [ "en" ], "reply": { "root": { "cid": "bafyreidlq6jjz5xrylrkkjp5kfc4qcacskoikf2nact455jfo5rf72rc7m", "uri": "at://did:plc:zrtxny34tclzxofwabv4ajol/app.bsky.feed.post/3lngir2ryas25" }, "parent": { "cid": "bafyreidlq6jjz5xrylrkkjp5kfc4qcacskoikf2nact455jfo5rf72rc7m", "uri": "at://did:plc:zrtxny34tclzxofwabv4ajol/app.bsky.feed.post/3lngir2ryas25" } }, "facets": [ { "index": { "byteEnd": 63, "byteStart": 15 }, "features": [ { "uri": "https://u-of-o-nmr-facility.blogspot.com/2009/11/probe-arcing.html", "$type": "app.bsky.richtext.facet#link" } ] } ], "createdAt": "2025-04-23T00:35:19.404Z" } }